Loading...
Thumbnail Image
Publication

Flowing afterglow apparatus for the study of ion-molecule reactions at high temperatures

Hierl, Peter M.
Viggiano, A. A.
Paulson, John F.
Morris, Robert A.
Mundis, Paul L.
Dale, Fred
Williamson, John S.
Seeley, John V.
Menendez Barreto, Melani
Dotan, I.
... show 1 more
Citations
Altmetric:
Abstract
We describe two versions of a high temperature flowing afterglow apparatus. With a stainless steel flow tube wrapped with heating tape we have obtained data over the range 300-1300 K. In a version with a ceramic flow tube in a commercial furnace we have obtained data over the range 300-1600 K. The ceramic version is designed to take data up to 1800 K, but we have encountered experimental problems at the upper temperature range. The design modifications to a standard flowing afterglow needed to make measurements at elevated temperatures are described in detail, as are problems associated with operating at elevated temperatures. Samples of data are given. (C) 1996 American Institute of Physics.
Description
Date
1996-06
Journal Title
Journal ISSN
Volume Title
Publisher
AMER INST PHYSICS
Research Projects
Organizational Units
Journal Issue
Keywords
Citation
Miller, TM; Dotan, I; MenendezBarreto, M; Seeley, JV; Williamson, JS; Dale, F; Mundis, PL; Morris, RA; Paulson, JF; Viggiano, AA; Hierl, PM. Flowing afterglow apparatus for the study of ion-molecule reactions at high temperatures. REVIEW OF SCIENTIFIC INSTRUMENTS. JUNE 1996. 67(6): 2142-2148.